Overview

The chemical dunk tank is specifically designed for use in BSL3 and BSL4 biosafety laboratories.

The dunk tank serves as a specialized pass-through system for transferring goods between sections of a cleanroom or biology laboratory, using liquid disinfection to ensure effective sterilization.

Based on a classic design, the PBSC dunk tank—also referred to as a High Containment Tank, Decontamination Tank, Hydrogen Peroxide Chamber, or Containment Barrier—features stainless steel lids mounted on durable cast stainless steel hinges. Inside the tank, a hinge-down submersion plate ensures items are fully immersed in the neutralizing chemical solution for effective decontamination.

The dunk tank features a manual locking mechanism, allowing the doors to provide a secure barrier seal even without the fluid in the tank.

It achieves a typical pressure loss of 80Pa at SAT over 30 minutes, starting from a pressure of 500Pa.

The chemical dunk tank can be equipped with a control panel offering PLC-controlled immersion times and integration with card swipe systems for enhanced access control.

For the safe transfer of items from hazardous environments, a recovery claw is included.

Custom designs are available—contact our technical department for more information and tailored solutions.

Effective Decontamination:

  • Ensures thorough immersion for comprehensive coverage.
  • Utilises chemical solutions to sterilize or neutralize contaminants on items.

PBSC offers a variety of hatch options tailored to meet your project requirements—click here to explore our range.

Downloads

Datasheet

Innovation. Design. Quality.

The dunk tank is expertly crafted from 5mm and 3mm stainless steel, featuring a center baffle, guide bars, an overflow/top-up pipe on one side, and durable locking stainless steel lids.

It is equipped with a stainless steel back plate that is securely bolted to the wall and grouted to the floor. Above the tank, a double-glazed vision panel with a viewing area of approximately 570mm x 600mm is included, along with rubber bump stops on both sides for added protection.

A stainless steel support frame with adjustable feet and a retrieval claw are provided as standard components, ensuring stability and functionality.

Lids:

  • Stainless steel 316
  • Phenolic Resin (High-Performance laminate)

 

Available in various sizes and configurations to suit specific operational needs

Following on-site installation, PBSC offers a Planned Preventative Maintenance and Servicing Plan with an extensive after-sales support service, including planned preventative maintenance programs, servicing, and spare parts.

 

 

READ MORE

Features

  • Mesh transfer system
  • Suitable for BSL3 and BSL4 containment barriers
  • Air tight performance
  • Air tight performance from the lids without the liquid in the tank
  • Design solutions to suit stud, block or cast concrete walls
  • Decontamination for small-scale items
  • A reliable solution

Applications

Appropriate for BSL3 and BSL4 containment barriers…

Biosafety dunk tanks are extensively used in pharmaceutical manufacturing and biotechnology, where strict cleanliness and environmental control are critical to safeguarding products and processes from contamination.

Technical Specification

  • Please contact PBSC for specific design options.
  • -
  • PBSC reserve the right to introduce changes in performance, configuration and technology, dimension, weights and materials in the course of technical progress. Images may not be accurate.
  • Some of the information/images/equipment described or shown may only be available at extra cost. Features & specifications will vary.

Options

  • Liquid Level Sensor
  • Internal light to aid visibility
  • Pneumatic interlock with PLC controls
  • Sink to side
  • Vision panel above unit
  • Clean side gassing ports
  • Design solutions to fit different types of concrete walls
  • Customized Designs